Recall of Trinica Ratchet Handles (included in the Anterior Lumbar Plate (ALP) Instrument Tray)

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Zimmer Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2013-RN-01092-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2013-10-24
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    The ratchet handles may not be effectively sterilised if left in their designated locations in the instrument tray. this is due to the current bracketing in the tray potentially interfering with complete steam penetration into the cannulated handle.
  • Action
    Zimmer is advising users to sterilise the Ratchet Handles externally to the instrument tray. Replacement trays are currently being designed and will be available by July 2014.

Device

  • Model / Serial
    Trinica Ratchet Handles (included in the Anterior Lumbar Plate (ALP) Instrument Tray) Part Number: 07.00438.001Instrument Tray Part Number: 07.01058.001ARTG Number: 179397
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “The safety of patients has always been, and continues to be, Zimmer Biomet’s top priority and it is our honor to be a leader in this industry for the past 90 years,” Zimmer Biomet told ICIJ in a statement. “We adhere to strict regulatory standards, and work closely with the FDA and all applicable regulatory agencies in each of our regions as part of our commitment to operating a first-rate quality management system across our global manufacturing network. The company added that it is focused on staying at the forefront of innovation and doing right by the millions of patients who rely on the company’s products.
  • Source
    DHTGA