Recall of STERIS 4085, 4085 SS and 4085 Feather Weight Leg Section (FLWS) Surgical tables

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Device Technologies Australia Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2014-RN-00884-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2014-08-08
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    Steris has determined there is a potential that if the surgical table feet become unlocked from the floor during a procedure, the hand control may incorrectly indicate that the table feet are locked and still contacting the floor. the user may not realise that the table is unlocked and is able to continue articulating the table using the hand control. no patient injury or adverse health outcomes have been reported due to this issue.
  • Action
    Should this occur, users are advised that the table can be relocked by pressing the Floor Lock button on the table hand control. As soon as parts are available a Device Technologies representative will contact customers to schedule the installation of new hardware and updated software on all affected Surgical tables. In addition, a new copy of the 4085 Surgical Table Operator Manual will be provided during this site visit. Users may continue to operate all 4085 series surgical tables prior to the completion of the update. This action has been closed-out on 19/08/2016.

Device

  • Model / Serial
    STERIS 4085, 4085 SS and 4085 Feather Weight Leg Section (FLWS) Surgical tablesProduct Codes: ST-01-360-1, ST-01-380-1 and ST-01-360-7All batches manufactured between 3 July 2009 and 20 June 2014Multiple affected serial numbersARTG number: 149156