Recall of SOMATOM Definition AS, Definition, Definition Edge, Definition Flash and SOMATOM Force

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Siemens Healthcare Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2017-RN-01355-1
  • Event Risk Class
    Class I
  • Event Initiated Date
    2017-11-08
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    Siemens has identified a potential risk of unnecessary radiation exposure due to a software issue in the care dose4d algorithm implemented in specific siemens ct scanners. this relates to possible incorrect tube current calculations by the care dose4d algorithm for head scans based on p.A. (posterior-anterior) or a.P. topograms. depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate. potentially, the care dose4d software will select the maximum tube current for the uppermost part of the skull, thus leading to unnecessary radiation exposure. the described issue will not occur when using a lateral topogram instead of a p.A. or an a.P. topogram. accordingly, siemens strongly recommends using topograms in lateral position for all head scans.
  • Action
    Siemens is advising users that the described issue will not occur when using a lateral topogram instead of a.p.a. or an a.p. topogram. Accordingly, Siemens strongly recommends using topograms in lateral position for all head scans. Siemens will be providing a software update as a permanent correction.