Recall of ROHO High Profile and Mid Profile Single Compartment Cushion with Sensor Ready Technology

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Permobil Australia.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2017-RN-00400-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2017-04-07
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    A problem was discovered internally with a supplied component for the affected products listed above. when a smart check by roho (smart check), model number smartcheck-01, is used to perform a smart setup with one of the affected cushions, the nonconforming part may adversely affect the setup process resulting in an improper cushion setup. sitting on an over inflated or under inflated sensor ready cushion may reduce or eliminate the benefits provided by the cushion, increasing the risk to skin or to other soft tissue.(please note that this risk is not present when the sensor ready cushion is used without the smart check nor is the risk present when the sensor ready cushion is used with the smart check and the setup is performed using the optional setup mode).
  • Action
    Permobil is advising users to return affected stock to their distributor. Distributors are requested to then contact Permobil to organise a replacement of the affected stock.

Device

  • Model / Serial
    ROHO High Profile and Mid Profile Single Compartment Cushion with Sensor Ready Technology Multiple Model NumbersSerial Numbers between A2920163 and A3185287Production dates from 8 January 2016 to 28 February 2017ARTG Number: 111106
  • Manufacturer

Manufacturer