Recall of Radiometer ABL90 Series analysers with software version 3.2 or below. An in vitro diagnostic medical device (IVD)

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Radiometer Pacific Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2017-RN-00714-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2017-06-07
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    In august 2015 radiometer advised users about a potential issue (tga ref. rc-2015-rn-00805-1) where a battery module overheated caused by the continued charging of a failing battery cell. the temporary countermeasure recommended, after the battery modules below 910-311p0009nxxx had been replaced, was for the capacity of the battery module to be checked at least every 12 months following the instructions that were provided.The manufacturer has now released an upgrade of the abl90 series software to software version 3.3 or higher. this new version monitors the charge level for the battery. if one or more failing battery cells are detected, the analyser status turns yellow and message 1559 “battery error detected” appears. if message 1559 has been active for four weeks in a row, the analyser enters operator intervention required with message 1560 “battery error still detected” and the analyser is locked until the battery has either been removed or exchanged.
  • Action
    Radiometer will contact users to arrange for the installation of the software upgrade.

Device

  • Model / Serial
    Radiometer ABL90 Series analysers with software version 3.2 or below. An in vitro diagnostic medical device (IVD)All Serial NumbersARTG Number: 228382