Recall of Puritan Bennett 840 Ventilator

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Covidien Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2014-RN-00618-1
  • Event Risk Class
    Class I
  • Event Initiated Date
    2014-06-05
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    Covidien has received customer reports of the loss of graphical user interface (gui) display information while the ventilator continues to provide breath support. customer reports describe the gui display screens (both upper and lower) becoming blank during patient use. this event interferes with the clinician's ability to view and/or change ventilator setting or clinician-set patient alarm parameters. loss of gui display due to a power supply unit (psu) failure is demonstrated by the ventilator entering a “gui inop” state i.E., if the breath delivery unit (bdu) loses communication with the gui, the bdu recognises this and enters a “gui inop” state. it initiates the appropriate alarm consisting of a visible alarm indicated by the illumination of the “gui inop” red light on the bdu and a high-priority audible alarm. the source of the psu voltage loss is a specific circuit responsible solely for the gui central processing unit voltages.
  • Action
    A Covidien representative will perform an inspection of affected ventilator(s) to determine if it has the affected component and if so, will replace defective PSU immediately. If users encounter a loss of display, they are advised to verify the patient’s respiratory and physiological stability ie, confirm the patient is receiving ventilatory support by observation of the expansion and contraction of the patient’s chest and assess current patient status by review of other monitoring (e.g., oxygen saturation, heart rate, blood pressure, etc). Also, users are advised to promptly transfer the patient to an alternate ventilator in a manner consistent with the institutional protocol and remove the affected ventilator from use until it has been serviced. This action has been closed-out on 02/02/2016.

Device

  • Model / Serial
    Puritan Bennett 840 VentilatorPower Supply Serial Numbers: 39302, 40221, 40355, 40465, 40324, 40327, 40450Vent Serial Numbers: 3510102854, 3510102868, 3510102839, 3510102853, 3510102849, 3510102848, 3510102852ARTG Number: 182309
  • Product Classification
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.
  • Source
    DHTGA