Recall of Philips HeartStart MRx cardiac monitor / defibrillators, models:M3535A, M3536A, M3536J, M3536M, M3536MC, M3536M1, M3536M2, M3536M3, M3536M4, M3536M5 and M3536M6 (manual defibrillator)

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Philips Electronics Australia Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2012-RN-00736-3
  • Event Risk Class
    Class I
  • Event Initiated Date
    2012-07-19
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    When heartstart mrx defibrillator monitors are used in external transport and ems environments the mechanical/electrical connection between the pads/paddles therapy cable (including pads cpr therapy cable and external paddles cable) and the mrx therapy connection port may experience higher than expected levels of stress causing accelerated wear.
  • Action
    Philips is revising the Instructions for Use (IFU) to include additional requirements and instructions for ongoing inspections of these devices, informing users that the lifespan of this device is 3 years and is providing support for users needing to replace cables.

Device

  • Model / Serial
    Philips HeartStart MRx cardiac monitor / defibrillators, models:M3535A, M3536A, M3536J, M3536M, M3536MC, M3536M1, M3536M2, M3536M3, M3536M4, M3536M5 and M3536M6 (manual defibrillator)All units manufactured prior to June 2012 (Model / manufacture date is printed on the primary label on the back of battery bay B)ARTG Number: 95661