Recall of NX workstation with NX 3.0.8950 image processing software

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by AGFA HealthCare Australia.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2017-RN-00131-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2017-01-27
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    Due to a memory issue, users are shown a warning message about unexpected error and possible incorrect data integrity. a reboot of the workstation computer after having received the warning message prevents the issue from occurring. at this stage, there has been one instance whereby the warning instructions were not followed and resulted in the image of one patient being archived in another patient’s record.
  • Action
    In order to prevent this issue, Agfa HealthCare is advising users to conduct a daily reboot of their workstation by completely turning off the computer and then restarting it. Users are further advised to follow all warning instructions as indicated on their NX workstation and reboot the workstation’s computer if a warning message appears. A new version of the NX software (3.0.9000) will be released shortly and users will receive this upgrade to their installations accordingly.