Recall of Multiple Models of GE Nuclear Medicine Systems:All VG and VG Hawkeye nuclear medicine systems, manufactured from 1996 - 2003: - Millennium VG - Millennium VG and Hawkeye- Discovery VH - VaricamAll Infinia nuclear medicine systems, manufactured from 2003 – 2014:- Infinia - Infinia and Hawkeye - Infinia and Hawkeye 4 All Brivo NM615, Discovery NM630, Optima NM/CT640 and Discovery NM/CT670 systems.

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by GE Healthcare Australia Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2014-RN-01052-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2014-09-26
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    Ge healthcare has recently performed field actions (tga ref.: rc-2013-rn-00629-1; rc-2013-rn-00679-1; rc-2013-rn-00695-1) on ge healthcare nuclear medicine system due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry . as part of this field action, ge healthcare is updating the preventative maintenance procedure and schedule and added an additional preventative maintenance check for loose fasteners on the nuclear medicine systems.
  • Action
    GE Healthcare is advising the customers to ensure that the organization that services the equipment uses the latest version of the Service manual and the Preventative Maintenance procedure available on GE website (link provided in the customer letter)

Device

  • Model / Serial
    Multiple Models of GE Nuclear Medicine Systems:All VG and VG Hawkeye nuclear medicine systems, manufactured from 1996 - 2003: - Millennium VG - Millennium VG and Hawkeye- Discovery VH - VaricamAll Infinia nuclear medicine systems, manufactured from 2003 – 2014:- Infinia - Infinia and Hawkeye - Infinia and Hawkeye 4 All Brivo NM615, Discovery NM630, Optima NM/CT640 and Discovery NM/CT670 systems.ARTG number: 200439
  • Manufacturer

Manufacturer