Recall of Molift Air 205 and 300 ceiling hoist (a powered lifting device)

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Patient Handling Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2016-RN-00975-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2016-08-25
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    The manufacturer has received a total of 4 reports where the bolt on the suspension coupling on molift air has become loose in the coupling. all reports came after technical testing and inspection. to date, no reports have been received regarding patient injury resulting from this issue. however, there is a potential risk for harm or injury if the bolt inside the coupling comes loose from the housing.
  • Action
    Patient Handling will arrange to upgrade the device with a retainer safety ring that will prevent the bolt within the suspension coupling coming loose. Qualified service personnel will install the upgrade kits on all affected systems. Consumers are advised to contact the place of purchase to organise a qualified service technician to install the upgrade kit. For more details, please see https://www.tga.gov.au/alert/molift-air-ceiling-hoist .

Device

  • Model / Serial
    Molift Air 205 and 300 ceiling hoist (a powered lifting device)Molift Air 205Model Numbers: 2501205, 2501210, 2501215, 2501220Molift Air 300Model Numbers: 2501300, 2501310, 2501315, 2501320Distributed between May 2013 and January 2016ARTG Number: 229696
  • Manufacturer

Manufacturer