Recall of Medtronic MiniMed Infusion Sets MiniMed Quick-set MiniMed mio MiniMed Silhouette MiniMed Sure-T

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Medtronic Australasia Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2017-RN-01182-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2017-09-11
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    Medtronic has become aware of recent reports of potential over-delivery of insulin shortly after an infusion set change. over-delivery of insulin can cause hypoglycaemia and in extreme cases, death. medtronic has received reports of hypoglycaemia requiring medical intervention potentially related to this issue. out of the millions of infusion sets sold each year, the reported incident rate requiring medical assistance related to this issue is less than 0.00005%. there have been no reports of death related to this issue.An investigation by medtronic has shown this issue can be caused by fluid blocking the infusion set membrane during the priming/fill-tubing process. a membrane blocked by fluid most likely occurs if insulin, alcohol, or water is spilled on the top of the reservoir or inside the tubing connector which then could prevent the infusion set from working properly.
  • Action
    For patients: 1. Users and HCPs with affected lot numbers (as confirmed with Medtronic using the dedicated webpage https://checklots.medtronicdiabetes.com or Global Help Line 1800 777 808) are to cease using the sets and, in the absence of unaffected alternatives, follow the steps contained in the supplied letter. Medtronic will replace the affected sets free of charge. 2. Cut out the bar codes on the boxes of recalled infusion sets (as per supplied picture) and return the bar codes in the reply paid envelopes that will be sent out with the new infusion sets. 3. Users are to discard, not return, the affected sets. 4. Customers are also requested to complete and sign the supplied Acknowledgment Form, and return as directed to Medtronic Australasia. For further information, please see https://www.tga.gov.au/alert/medtronic-minimed-infusion-sets

Device

  • Model / Serial
    Medtronic MiniMed Infusion SetsMiniMed Quick-setMiniMed mioMiniMed SilhouetteMiniMed Sure-TARTG Number: 119216
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.
  • Source
    DHTGA