Recall of Medtronic CryoConsole (A unit that applies a gaseous or liquid refrigerant to withdraw heat from target tissues or indirectly through contact with a cryogen-cooled probe.)

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Medtronic Australasia Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2015-RN-00853-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2015-09-04
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    Medtronic has identified a performance issue with a usb memory component that contains some of the system's software within a subset of cryoconsole models. testing indicated that the write speed of the usb memory component from a single supplier lot is slower than those compared to other lots. based on data from field complaints and internal testing, the difference in speed is known to cause the following issues:-1. during inflation of a cryoballon catheter, false detection of balloon deflation and display of system notice 22406 (the balloon is deflated);2. during a procedure, display of system notice 50008 (the system has detected a software error and stopped the injection).No injuries have been reported.
  • Action
    Users are advised that if a system notice appears, to carefully read and follow the instructions on the screen of the console. A Medtronic representative will also contact users to arrange a service of their CryoConsole system. This action has been closed-out on 18/08/2016.

Device

  • Model / Serial
    Medtronic CryoConsole (A unit that applies a gaseous or liquid refrigerant to withdraw heat from target tissues or indirectly through contact with a cryogen-cooled probe.)Models 106A3, 106E2, 106A2-K ARTG Number: 164650
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.