Recall of James Leckey Design Mygo Stander

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Otto Bock Australia Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2016-RN-00854-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2016-06-29
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    It has been identified that the initial design of the angle adjustment mechanism for the stander allowed the user to lock out the safety pull pin on the frame and adjust the angle of the stander using one hand. if the operator’s hand slipped off the turning handle when a patient was in the frame then the handle could “self-wind” and spin rapidly as the stander returns to the upright position. whilst there is no risk to the patient in the frame, the fast spinning handle does pose a risk of injury to the operator if they try to take hold of the handle.These issues only impact devices manufactured between january 2012 and august 2013 as the design of the pull pin safety mechanism was changed in august 2013. it is no longer possible to lock the pull pin out and two hands are now required to adjust the stander angle.
  • Action
    Otto Bock is arranging for distributors to contact users of the affected devices and arrange for replacement of the pull-pin on the device with one that cannot be locked by the operator. The replacement safety pull pin will require the user to hold the pull pin with one hand while turning the adjustment handle with the other hand. If their hand slips off the turning handle, they can simply release the pull pin and the handle will stop winding. For more information, please see https://www.tga.gov.au/alert/james-leckey-design-mygo-stander .