Recall of Jackson-Pratt Perforated and Hemaduct Wound Drainage Systems

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Medline International Two Australia Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2017-RN-00565-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2017-05-03
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    Medline is recalling specific jackson-pratt perforated and hemaduct wound drainage systems, due to a potential breach of sterility in the packaging. in a small percentage of product (approximately 1.6%), the inner packaging (polybag) is caught in the seal area of the outer tyvek pouch, potentially compromising the sterility of the package contents. use of impacted products could result in an increased risk of infection. to date, the manufacturer has not received any complaints or reports of infection associated with this issue.
  • Action
    Medline is advising customers to remove affected codes/lots from their stock holdings and to also segregate affected units for Medline to collect. Medline will organise pick-up of the affected stock and a credit note will be issued.

Device

  • Model / Serial
    Jackson-Pratt Perforated and Hemaduct Wound Drainage SystemsProduct Codes: JP-HUR101, SU130-1308, SU130-1309, SU130-1310, SU130-1311 and SU130-1411 Batch Numbers: 1161468, 1161228, 1161396, 1161413, 1161460, 1161461, 1161230, 1161399, 1161247, 1161251, 1161622, 1161623 and 1161253ARTG Number: 169567
  • Manufacturer

Manufacturer