Recall of Invacare TDX series and Storm Series power wheelchairs with 20”-24” width configurations with Formula CG Recline seating system (i.e. Tilt/Recline, Tilt/Recline/Elevate, or Recline only) and a conventional style back

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Invacare Australia Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2015-RN-00356-1
  • Event Risk Class
    Class I
  • Event Initiated Date
    2015-04-24
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    Invacare has recently identified a potential pinch point affecting the wiring of the listed power wheel chairs. there is a possibility that if the slack in the wires is not routed and secured correctly, and with flexing of the back pan under the user’s weight, the headrest knob or clamp bracket may pinch, damage or cut the wire creating a hazard. smoking, sparking, burning or fire may occur if any wire in this area is repetitively pinched, at the same location, over a period of time. not all impacted chairs are expected to have slack in the wiring that is not routed and secured correctly. invacare has received one complaint and there have not been any reported events involving injury associated with this issue.
  • Action
    Dealers are advised to identify wheelchair owners and notify them of this issue and proposed actions using the consumer letters provided by Invacare. Dealers are also advised to contact their respective customers to schedule an appointment to inspect the potentially impacted power wheelchair as soon as possible. A visual inspection of the wiring shall be conducted for each impacted chair and where required the wiring will be (corrected) routed and secured in accordance with the Power Wheelchair Wiring Guide provided to dealers. For more details, please see http://www.tga.gov.au/alert/invacare-tdx-and-storm-series-power-wheelchairs . This action has been closed-out on 18/08/2016.

Device

  • Model / Serial
    Invacare TDX series and Storm Series power wheelchairs with 20”-24” width configurations with Formula CG Recline seating system (i.e. Tilt/Recline, Tilt/Recline/Elevate, or Recline only) and a conventional style backModels:TDX Series: TDXSP, TDXSP-AUS, TDX SR-AUS & TDX SRStorm Series: 3GTQ-MCG, STORM TORQUE, STORM TORQUE-AUS, RANGERX, 3GRX-CG, ARROW 3G & 3GAR-TSMultiple serial numbersManufactured from October 1, 2006 - July 31, 2014ARTG number: 160385
  • Product Classification
  • Manufacturer

Manufacturer