Recall of Integra Camino Intracranial Pressure (ICP) Monitor and the Integra Licox PtO2 and Temperature MonitorIntegra Camino Intracranial Pressure (ICP) Monitor

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Integra Neurosciences Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2016-RN-00030-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2016-01-08
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    The speakers in camino icp monitors (cam02) and the licox pto2 and temperature monitors (lcx02) may potentially malfunction. if this occurs, the monitors will stop providing an audible tone at start-up or at alarm points during use. although unlikely, this may lead to central herniation that could potentially result in brain death or reversible ischemia; moderate hypoxic brain injury; elevation of icp or decrease in pto2. no reports of patient injury or adverse events have been received.
  • Action
    The affected units are to be returned to the manufacturer for a software upgrade to lower the input signal supplied to the speaker assembly. The monitor case will also be modified to increase the decibel level.

Device

  • Model / Serial
    Integra Camino Intracranial Pressure (ICP) Monitor and the Integra Licox PtO2 and Temperature MonitorIntegra Camino Intracranial Pressure (ICP) MonitorCatalogue Number: CAM02 All serial numbers affectedIntegra Licox PtO2 and Temperature MonitorCatalogue Number: LCX02All serial numbers affectedARTG Number: 216915 and 217355
  • Manufacturer

Manufacturer