Recall of Innova 3100, Innova 3100IQ and Innova 2100IQ Cardiovascular X-ray Imaging Systems

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by GE Healthcare Australia Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2014-RN-00018-1
  • Event Risk Class
    Class I
  • Event Initiated Date
    2014-01-09
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    The innova system can unexpectedly stop delivering x-rays after a power-on or a reset cycle which may result in the loss of real-time interventional imaging. no injuries have been reported due to this issue.
  • Action
    GE Healthcare is advising users to ensure the Innova system emits x-rays before starting any procedure by completing a quick system functional check. If the system displays an "X-ray not available" error message, GE is advising users to reset the system. GE Healthcare will be correcting all affected systems.