Recall of HS66 Turbo LTSF Steriliser

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Getinge Australia Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2014-RN-00908-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2014-08-18
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    The problem identified indicates a leakage at a critical valve due to clogging of solidified formalin solution, also known as paraformaldehyde. this leakage might affect the efficacy of the ltsf sterilisation cycle and cause potential non-sterility. the problem appears stealthy until clogging has reached the level where the built-in leakage monitoring system of the steriliser is able to indicate leakage and cause an alarm. the potential risk caused by clogging is that sterile filtered air can be introduced into the sterilizer chamber during the sterilization cycle and may partially block penetration of formaldehyde at items subjected to sterilization. the risk for non-sterility is assessed to increase at hollow instruments compared to solid instruments.
  • Action
    A Getinge representative will schedule a service technician visit to inspect, clean and take preventive measures at your sterilizer for the problem not to reappear. In the meantime all users are advised to take precautions as follows: - Depending of the occurrence and frequency of LTSF sterilization at your sterilization department use biological indicators for monitoring the sterilization efficacy at a daily or weekly basis; - When growth is indicated by biological indicators - stop using the sterilizer and contact your Getinge representative for prompt assistance. This action has been closed-out on 03/06/2016.