Recall of Homechoice Pro APD System (peritoneal dialyser)

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Baxter Healthcare Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2013-RN-01117-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2013-11-01
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    An occluder blade clamps the solution lines if a power failure occurs or when the cycler is off preventing uncontrolled solution delivery to the patient. the occluder may contain sharp edges and/or separate from the back plate, which may cause the occluder assembly to fail resulting in an unrecoverable slowflow/no flow alarm/system error.
  • Action
    Baxter is informing the affected hospitals of the issue. Affected devices are to be returned and replaced with unaffected devices. Baxter is advising patients if there is an unrecoverable failure to perform manual peritoneal dialysis to continue therapy.

Device

  • Model / Serial
    Homechoice Pro APD System (peritoneal dialyser)Product Code: R5C8320Serial Number: 300905, 300882, 300928, 300889ARTG Number: 172780