Recall of HeartWare Ventricular Assist System batteries

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Heartware Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2014-RN-00902-1
  • Event Risk Class
    Class I
  • Event Initiated Date
    2014-08-14
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    In may 2014, heartware distributed recall for product correction letter (tga ref: rc-2014-rn-00510-1/ heartware ref: fsca apr2014.1) following an increase in reported power management complaints related to earlier than expected battery depletion. information on early battery depletion, identification of potentially faulty batteries, and recommended practices for power management were provided in the recall for product correction letter.Heartware is expanding the action to include the recall of batteries produced within the serial number ranges: bat000001 to bat039999 and bat090000 to bat099999. batteries produced in these ranges are more likely to exhibit premature or unrecognized deterioration of battery capacity, which poses a risk to the patient since it may result in a pump stop.
  • Action
    HeartWare is requesting their clinicians to review their hospital’s current inventory and remove and quarantine any affected batteries. At each patient’s next regularly scheduled visit, inspect their batteries and replace affected batteries with batteries having serial numbers BAT040000 to BAT089999 or BAT100000 and above. If a patient does not have a regular visit scheduled within the next 90 days, contact that patient by telephone to check the serial numbers of the patient’s batteries. This action has been closed-out on 13/10/2016.

Device

  • Model / Serial
    HeartWare Ventricular Assist System batteriesProduct code: 1650Serial number range for the affected batteries: BAT000001 to BAT039999; BAT090000 to BAT099999ARTG number:181875
  • Product Classification
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.
  • Source
    DHTGA