Recall of Halo Flex Energy Generators (Endotherapy electrosurgical unit)

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Covidien Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2012-RN-00824-3
  • Event Risk Class
    Class II
  • Event Initiated Date
    2012-08-14
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    Covidien has received reports related to a generator malfunction where the haloflex energy generator fails to enter “standby mode” when initially powered-on. the generator became stuck in the self-test mode upon initial power-on, did not enter standby mode, and therefore could not be used for patient care.
  • Action
    The customer letter provides temporary work around instructions. Covidien is developing a software revision, which will be available by the end of August 2012 that will prevent the malfunction from recurring.

Device

  • Model / Serial
    Halo Flex Energy Generators (Endotherapy electrosurgical unit)Model Number: 1190A-230AAll serial numbers ARTG Number: 140684
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.