Recall of H LED and PowerLED Ambient Light Modules (Surgical light)

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Maquet Australia Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2014-RN-00347-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2014-03-20
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    There is a potential risk of failure concerning the fixation of the ambient light module. it has been observed by some of the customers that one or several fixing tabs were broken, which led the module to detach, remaining attached by its wires only. the occurrence rate observed is lower than 1% and is mainly seen in the early use of the devices.
  • Action
    All customers are asked to check if the ambient light module is correctly fixed. The procedure for this is provided with the customer letter. If any issues are detected by the customer, precautions should be taken in order to avoid the component falling and a Maquet representative should be contacted immediately in order to replace the broken ambient light module with a new one.

Device

  • Model / Serial
    H LED and PowerLED Ambient Light Modules (Surgical light)Model numbers: PWD700 and H LED 700 All lots manufactured from December 2006 to December 2012ARTG Number: 162037