Recall of Freelite Human Lambda Free kit for use on SPAPLUS

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by In Vitro Technologies Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2016-RN-01425-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2016-11-09
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    The manufacturer has advised that investigations of the products have identified a change in the calibration curve shape that has occurred since product release. increased activity in the calibration curve between calibrators 5 and 6 has been observed and they have received reports of increased false prozone flags which is consistent with this issue.
  • Action
    Customers are advised to immediately quarantine and dispose of any affected lots held by the customers and also to remove any affected lots from the SPAPLUS instrument. It is also advised to discuss the issues with the Medical Director to identify any further actions are required including reviewing the previously reported results.

Device

  • Model / Serial
    Freelite Human Lambda Free kit for use on SPAPLUSItem Numbers: TBSLK018S, TBSLK01810SLot Numbers: 400012, 400257Expiry: Nov 2017
  • Manufacturer

Manufacturer