Recall of Frazier suction handle

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by ConvaTec Australia Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2018-RN-00061-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2018-01-24
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    Convatec has been made aware of overseas complaints from users regarding swivelling of the suction tip of the handle of these devices. depending on the type of procedure, there is a potential for serious harm such as delayed recovery/delay in treatment, and tissue damage. it is very likely that the user may still be able to continue suctioning even after the suction tip starts to swivel within the handle. the chances of detecting the swivel by the user remain high.No complaints for this issue have been received to date in australia. no patient harm has been reported as a result of this issue.
  • Action
    Convatec is advising users to inspect the Frazier suction handle for a loose tip prior to use. Additional/alternative Frazier units may be kept on standby just in case users experience swivelling of the suction tip. The manufacturer is currently validating a revised manufacturing process after shifting the bending fulcrum of the bending fixture from the handle to the metal tube thus reducing the impact of the bending force on the handle. After successful validation and product testing, the revised product will be introduced to the market.

Device

  • Model / Serial
    Frazier suction handleProduct code: UM1912 (12 Fr)Lot Number: CGCA06-01ARTG Number - 289102 (ConvaTec Australia Pty Ltd - Suction tip, general-purpose, single use)
  • Manufacturer

Manufacturer