Recall of Fortify, Fortify Assura, Quadra Assura, Quadra Assura MP, Unify, Unify Assura and Unify Quadra devices(Implantable cardiac defibrillator (ICD) and cardiac resynchronisation therapy defibrillator (CRT-D) devices) Manufactured between January 2010 and May 2015 Fortify VRFortify ST VRFortify Assura VRFortify DRFortify ST DRFortify Assura DRUnifyUnify QuadraUnify Quadra MPUnify AssuraQuadra AssuraQuadra Assura MP

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by St Jude Medical Australia Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2017-RN-01122-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2017-08-28
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    St jude medical (sjm) is providing additional information following on from the previous notification in october 2016 (rc-2016-rn-01316-1).Previously there was no method to identify devices subject to premature battery depletion prior to reaching the eri voltage. sjm have developed battery performance alert (bpa) for patients who have impacted devices.The bpa will provide notification of abnormal battery performance prior to premature battery depletion, & before eri voltage. battery information is uploaded nightly to merlin.Net where the most recent 32 days of data are analysed each day to determine any anomalous battery voltage trend. once the bpa is triggered, notification is provided to physicians through merlin.Net and the merlin programmer during routine follow-up evaluations. for patients not followed remotely with merlin.Net, the status of their battery/whether the bpa has triggered an alert can only be determined with in-person interrogation using the merlin programmer.
  • Action
    St Jude Medical is advising customers: 1. Do not implant unused affected devices. 2. Conduct patient follow-up per standard practice. 3. Prophylactic device replacement is NOT recommended because complications following replacement have been reported to occur at a greater rate than the rate of harm associated with premature battery depletion due to lithium cluster induced shorts. 4. In the event of a BPA or ERI indicator in these devices, immediate device change is recommended. (Updated recommendation) 5. Follow directions in the customer letter regarding battery monitoring in patients of already-implanted devices.

Device

  • Model / Serial
    Fortify, Fortify Assura, Quadra Assura, Quadra Assura MP, Unify, Unify Assura and Unify Quadra devices(Implantable cardiac defibrillator (ICD) and cardiac resynchronisation therapy defibrillator (CRT-D) devices)Manufactured between January 2010 and May 2015Fortify VRFortify ST VRFortify Assura VRFortify DRFortify ST DRFortify Assura DRUnifyUnify QuadraUnify Quadra MPUnify AssuraQuadra AssuraQuadra Assura MPMultiple ModelsMultiple ARTG Numbers
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “We are in constant communication with regulatory agencies and competent authorities worldwide which allows us to implement global recalls or in-country communication quickly and effectively,” Abbott, which now owns St. Jude Medical told ICIJ in a statement. In addition to sending global notices to physicians worldwide, we also make sure that product advisories are available online and classification of product recalls and product advisories are determined by global regulatory bodies which can impact the timing in any given country. MD companies follow varying regulations in different countries. In come countries software is not regulated so a recall in one country related to software would not be classified as a recall or field action in another. In addition, review cycles within the regulatory process can be different in each country which can impact communication and recall timing.
  • Source
    DHTGA