Recall of FOLFusor SV 0.5 ml/hr

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Baxter Healthcare Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2014-RN-00447-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2014-04-17
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    The direction insert for the folfusor sv 0.5 ml/hr states that infusion rates may be up to 10% greater than the nominal infusion rate of 0.5 ml/hr. flow rate testing indicates that infusion rates may be higher and could result in a 15% greater than nominal infusion rate.
  • Action
    Baxter is notifying their customers that infusion rates may be higher and could result in a 15% greater than nominal infusion rate rather than 10% as stated in the IFU. Baxter is also reiterating the factors that can contribute to over infusion. These factors along with the 15% flow rate accuracy could result in infusion rates in excess of 30% greater than the nominal flow-rate. Baxter is currently investigating options to bring the flow rate into specification.

Device

  • Model / Serial
    FOLFusor SV 0.5 ml/hrProduct Code: 2C4700KAll lot numbers up to and including 14B039ARTG Number: 168893