Recall of Fetch 2 Aspiration Catheter (used for the removal of fresh, soft emboli and thrombi from vessels in the peripheral and coronary vasculature)

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Boston Scientific Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2016-RN-00359-1
  • Event Risk Class
    Class I
  • Event Initiated Date
    2016-03-24
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    Boston scientific is implementing a recall for all models of the fetch 2 aspiration catheter as a result of receiving reports from the field of catheter shaft breaks; including cases requiring the use of a snare or other intervention to retrieve the broken shaft from the patient. no serious patient injuries have been reported.
  • Action
    Boston Scientific is advising users to quarantine affected stock and return for credit. This action has been closed-out on 27/01/2017.

Device

  • Model / Serial
    Fetch 2 Aspiration Catheter (used for the removal of fresh, soft emboli and thrombi from vessels in the peripheral and coronary vasculature)Material Numbers (UPN): 109400-002, 109400-003, 109400-004 AND 109400-005Multiple batch numbers (see attached)Batch expiration date range: March 2016 to June 2017ARTG number: 181630
  • Product Classification
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “We take a patient-first approach to assessing the applicability of every recall and communicate to regulatory bodies in all geographies where the recalled device is sold,” Boston Scientific said in a statement to ICIJ. “We have coordinated several recalls across many countries in a timely manner,” the company said, adding that it complies with all national laws, which can often vary and require different processes for reporting information or taking action on recalls. The company said it uses a rigorous and uniform process to take action on recalls and that “when we initiate a field action (e.g. recall, safety alert), every customer who has received an affected product receives a communication that includes a letter for the physician.”
  • Source
    DHTGA