Recall of Expression Information Portal - IP5 (secondary display control unit for Expression MRI patient monitoring system)

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Philips Electronics Australia Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2014-RN-00349-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2014-03-20
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    Continuous use of the ip5 for more than 24 days, without power cycling, results in a graphical user interface (display screen) error, which causes the numeric data on the display to freeze and disables some menu keys. waveform data displayed and audible alarms are not impacted by this problem and continue to function properly.
  • Action
    Philips is advising end users to ensure that the IP5 is power cycled every 24 hours. A correction is being investigated and will be provided once available.