Recall of Epix and Direct Drive Laparoscopic Grasper

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Applied Medical Australia Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2013-RN-00434-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2013-05-13
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    If a great enough force is applied when the ratchet trigger is activated, there is a potential for the trigger to fracture and the jaws to remain in the closed position.
  • Action
    Applied Medical is requesting their customers to inspect their stock and return any affected units.

Device

  • Model / Serial
    Epix and Direct Drive Laparoscopic GrasperModel numbers: C4130 and C4140Multiple lot numbers affected
  • Manufacturer

Manufacturer