Recall of Emergency Room Examination Table

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Getinge Australia Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2018-RN-00653-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2018-06-05
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    Getinge have received reports where the screws holding the reclining surface on the frame have loosened or completely fallen out. the screws that hold the reclining surface on the frame are not secured against unintentional loosening and the tables may experience this issue with daily use over a number of years. if all screws loosen / become detached and this goes unnoticed, the reclining surface of the table may detach resulting in injury to a patient or operator as a result of falling. to date, no reports of patient injury have been received relating to this issue.
  • Action
    Getinge are advising they will be in contact with impacted users to replace the screws on the table with new screws secured with detent edge discs.

Device

  • Model / Serial
    Emergency Room Examination Table Models: 4203.9602 ARTG number: 163023(Getinge Australia - Table, operation)