Recall of Eluvia 150mm and Innova 200mm Stent Systems

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Boston Scientific Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2017-RN-01391-1
  • Event Risk Class
    Class I
  • Event Initiated Date
    2017-11-07
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    Boston scientific is initiating a removal from the market of the150mm eluvia drug-eluting vascular stent system and the 200mm innova self-expanding stent system due to elevated complaint rates for partial stent deployment.There have been 190 complaints of partial deployment. partial deployment occurs when the stent is unable to be fully released from the delivery system. part of the stent can become anchored in the vessel while the rest of the stent remains within the delivery system. the most common reported injury has been additional medical or minor surgical intervention, vessel trauma or prolongation of the implant procedure. however, boston scientific has received reports in 14 cases of major surgery to retrieve the stent/delivery system or to correct vascular compromise.
  • Action
    Boston Scientific is advising users to inspect stock and quarantine the affected products for return for credit.

Device

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “We take a patient-first approach to assessing the applicability of every recall and communicate to regulatory bodies in all geographies where the recalled device is sold,” Boston Scientific said in a statement to ICIJ. “We have coordinated several recalls across many countries in a timely manner,” the company said, adding that it complies with all national laws, which can often vary and require different processes for reporting information or taking action on recalls. The company said it uses a rigorous and uniform process to take action on recalls and that “when we initiate a field action (e.g. recall, safety alert), every customer who has received an affected product receives a communication that includes a letter for the physician.”
  • Source
    DHTGA