Recall of EliA Sample Diluent

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Abacus ALS Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2017-RN-00348-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2017-03-10
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    There have been reports of instrument malfunctions for phadia 250 due to deformed bottles in the above-mentioned products and lots. the deformation of the bottle can cause an erroneous volume detection by the instrument, which will result in insufficient aspiration of sample diluent, thus generating insufficient dilution of patient samples. the deformation is described as a flange or brim, located on the inside lower part of the vial opening. the instrument malfunction due to deformed bottles could cause erroneous test results, mainly a false positive, with regard to all elia assays. the frequency of this bottle defect is very low and not every defective bottle will cause the error. an instrument malfunction caused by deformed bottles will generate the instrument error message “3-145 right arm liquid detection below lower limit with target 2”.
  • Action
    Users are notified to stop using the affected lots and record the lot numbers and quantity of any unused products prior to discarding them. Abacus ALS will organise replacements at no charge. Also, users are advised to discuss the contents of the letter with the Medical Director to ascertain if any further actions are required.

Device

  • Model / Serial
    EliA Sample DiluentArt. No. 83-1023-01Lot Numbers: JM3Y, JS0B, JP87, JK5N, JU9CBottle Lots: BVXDR/JM3W, BVXDT/JS0A, BVXDS/JP83, BVXDP/JK5M, BVXDU/JU9BARTG Number: 229543