Recall of DxH 500 Open-Vial Haematology System, with software version 1.0.2 and 1.1. An in vitro diagnostic medical device (IVD)

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Beckman Coulter Australia Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2017-RN-01401-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2017-11-08
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    An internal investigation has identified that if the bleach cycle is interrupted or aborted, the dxh 500 does not provide any warning message to the operator and the system does not sufficiently flush out residual bleach prior to sample analysis. should a bleach cycle get interrupted, the subsequent software recovery to perform a diluter reset may not be sufficient to prevent the first immediate sample run from being affected. this may result in incorrect patient results.There have been no reports of injuries as a result of this issue.
  • Action
    Beckman Coulter is advising users that if the bleach cycle is interrupted or aborted to ensure the daily shutdown and daily checks are performed prior to running any patient samples. Beckman Coulter will be performing a software update as a permanent correction. A review of previously generated results is at the discretion of the laboratory director.

Device

  • Model / Serial
    DxH 500 Open-Vial Haematology System, with software version 1.0.2 and 1.1. An in vitro diagnostic medical device (IVD)Reference Number: B40601ARTG Number 208574
  • Manufacturer

Manufacturer