Recall of Discovery MR450, MR750, MR750w and Optima MR450w products, running software versions DV22.0_V02, DV22.1_V01, DV23.0_V01 (MRI System)

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by GE Healthcare Australia Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2013-RN-00376-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2013-04-29
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    The mr product can generate image headers with non-unique series identifiers, which when exported to some pacs systems, results in the series being associated with the incorrect study on the pacs. the intended study will not have the correct series listed and an unintended study will have it listed instead. in all cases, patient identifying annotation remains correct.
  • Action
    GE Healthcare is providing temporary work around instructions and implementing a software update.

Device

  • Model / Serial
    Discovery MR450, MR750, MR750w and Optima MR450w products, running software versions DV22.0_V02, DV22.1_V01, DV23.0_V01 (MRI System)ARTG Number: 193173