Recall of Dimension Creatinine and Dimension Vista Creatinine Assays

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Siemens Healthcare Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2015-RN-00680-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2015-07-28
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    Siemens healthcare diagnostics has determined that dimension creatinine (crea) may produce depressed results up to 12.7% (or approximately 18 µmol/l) for a creatinine result at 133 µmol/l when haemolysis exceeds 300 mg/dl [0.19 mmol/l] of haemoglobin. dimension vista creatinine (crea) may produce depressed results up to 22% (or approximately 27 ìmol/l) for a creatinine result at 133 ìmol/l when haemolysis exceeds 200 mg/dl [0.12 mmol/l] of haemoglobin.Depressed creatinine results due to haemolysis would not be expected to significantly impact medical decisions when using this assay. the potential for injury is remote due to continued monitoring, additional diagnostic testing and correlation to patient history and presentation.
  • Action
    A workaround is provided in the form of revised interference information via the Customer Letter.

Device

  • Model / Serial
    Dimension Creatinine and Dimension Vista Creatinine AssaysDimension CREA AssayCatalogue Number: DF33A Siemens Material Number (SMN): 10444968Dimension Vista CREA AssayCatalogue Number: K1033 Siemens Material Number (SMN): 10445161All in-date and future lots are affectedARTG number: 181689