Recall of Dimension Clinical Chemistry Systems HB1C Flex Reagent Cartridge (An in vitro diagnostic medical device (IVD))

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Siemens Ltd Diagnostics Division.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2013-RN-00316-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2013-04-15
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    Issue 1 - siemens healthcare diagnostics have confirmed a positive bias for values below 6%, there is a positive bias of 0.3% with an upper 95% confidence interval of 0.7%.The issue may impact quality control (qc) materials, patient samples, and proficiency survey samples. issue 2 - siemens healthcare diagnostics have confirmed a higher frequency of "above assay range" flags. there is a higher frequency of flags is related to higher recoveries of sample haemoglobin (hb) values (hb >25g/dl or >15.5mmol/l). the hb value is used to calculate the final output hba1c ratio (% hba1c [mmol/mol] result. results are suppressed (not reported) with the "above assay range" flag.
  • Action
    Siemens is asking users to destroy affected lots and will replace any recalled units.

Device

  • Model / Serial
    Dimension Clinical Chemistry Systems HB1C Flex Reagent Cartridge (An in vitro diagnostic medical device (IVD))Catalogue Number: DF105ASiemens Material Number (SMN): 10483822Lot Number: GA3009, GA3141, GA3197, GA3247, GA3113, GA3162, GB3211, GA3267, GA3134, GA3169, GA3232, GB3281, GC3302ARTG number: 181689