Recall of DePuy Synthes Trauma DHS/DCS Impactor TipInsert for DHS/DCS Impactor No. 338.280, Single

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Johnson & Johnson Medical Pty Ltd t/a DePuy Synthes.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2015-RN-00184-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2015-03-05
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    The dhs/dcs impactor is composed of two parts: a replaceable plastic tip which is the insert for dhs/dcs impactor (338.260), and a metallic shaft (338.280). no biocompatibility testing has ever been completed by the manufacturer, synthes for the material (polyamide 6.6 with 20% carbon fibre reinforcement, tecamid 66 cf 20) for the insert and no evidence was found that it has been completed by other entities. potential for harm exists due to the potential for device breakage and may include surgical delay and adverse tissue reaction.
  • Action
    Johnson & Johnson Medical is replacing the plastic tip of the insert with a metal tip. Once the redesigned tip is available Johnson & Johnson Medical will arrange for the recovery and replacement of the old impactor tip. In the interim to ensure the impactor is available for emergency surgery, Johnson & Johnson Medical has provided end users with inspection instructions to check for any wear and tear. Alternatively, hospitals can return the tip for a credit note. This action has been closed out on 14 June 2017.