Recall of da Vinci Si and Xi Surgical System - Touchpad Compact Flashcardsda Vinci Si/Xi SSC Touchpad

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Device Technologies Australia Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2016-RN-00312-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2016-03-22
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    Intuitive surgical has found that compact flash cards used in touchpads on specific da vinci si and xi surgical systems to save user preferences, have the potential to become corrupted when the touchpads experience an abrupt power loss and/or are powered on and off within a short period of time. the corruption of the compact flash occurs when power is cycled. if the compact flash has been corrupted, the da vinci system will display a fault during the start-up of the system. in the worst case scenario, the da vinci surgical system will encounter a non-recoverable fault, preventing use of the system. the error typically presents itself during start-up, so will not likely occur during a procedure (unless a mid-procedure power restart is required).
  • Action
    Device Technologies is advising users that a representative will be replacing the affected compact flash cards with unaffected products. This may require users to re-enter their user preferences. This action has been closed-out on 27/01/2017.

Device

  • Model / Serial
    da Vinci Si and Xi Surgical System - Touchpad Compact Flashcardsda Vinci Si/Xi SSC TouchpadPart Number:187505-04 da Vinci Xi PSC TouchpadPart Number: 187716-03 ARTG Number: 97348
  • Manufacturer

Manufacturer