Recall of Covidien Endo GIA Articulating Reloads with Tri-Staple Technology

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Medtronic Australasia Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2018-RN-00655-1
  • Event Risk Class
    Class I
  • Event Initiated Date
    2018-06-06
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    Medtronic has identified the potential for the device to have a misassembled (reversed) sled component. the sled component is responsible for staple deployment. when the device is applied to tissue and fired, a misassembled sled may result in the tissue being transected without the deployment of staples to seal the point of transection. this may result in bleeding, immediate anastomotic leak, and/or the potential for death if this is used on a critical vascular structure such as the pulmonary artery and renal artery. internationally, medtronic have received one report of injury related to this issue. to date, there have been no reported injuries in australia.
  • Action
    Medtronic is advising customers to immediately discontinue use of the affected products. Unused product should be quarantined and returned to Medtronic by following the instructions outlined on the Customer Letter. Medtronic will arrange collection of stock from customers and issue a credit for returned goods.

Device

  • Model / Serial
    Covidien Endo GIA Articulating Reloads with Tri-Staple TechnologyEndo GIA 45mm Articulating Medium/Thick Reload with Tri-Staple Technology Item Code: EGIA45AMT Lot: N7L0521KXARTG:181429 Endo GIA 45mm Extra Thick Black Articulating Reload with Tri-Staple TechnologyItem Code: EGIA45AXT Lots: N8A0907KX and N8A0146KX ARTG:181429Endo GIA 60mm Articulating Vascular/Medium Loading Unit with Tri-Staple TechnologyItem Code: EGIA60AVM Lot: N8B0088KX ARTG:186247
  • Product Classification
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.
  • Source
    DHTGA