Recall of Chorion Villus Biopsy Needle Set

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by William A Cook Australia Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2018-RN-00835-1
  • Event Risk Class
    Class III
  • Event Initiated Date
    2018-06-26
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    Cook medical has identified the diagram on the product label of the chorion villus biopsy needle set (k-cvns-1821-robinson-et) is incorrect. it shows that the 18ga needle has an echotip, and the 21ga needle does not, whereas the product is designed such that the 21ga needle has the echotip and the 18ga needle does not.The tip of the larger gauge guide needle is likely to be visible on the ultrasound regardless of a non-echogenic tip. in the event that the labelling error resulted in a clinician expecting the echogenic tip to be present on the guide needle, instead of the sampling needle, the clinician may not be able to adequately identify the tip of the inserted access needle.To date, there have been no reports of harm relating to this issue.
  • Action
    William A Cook is advising customers to inspect stock and return affected product.

Device

  • Model / Serial
    Chorion Villus Biopsy Needle SetCatalogue Identifier: K-CVNS-1821-ROBINSON-ETOrder Number: G26661All lots manufactured up to 4 June 2018ARTG Number: 138806(William A Cook Australia Pty Ltd - Chorionic Villi Biopsy Needle Set - Needle, biopsy, single use)
  • Manufacturer

Manufacturer