Recall of Charcot Osteotome Chisels

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Zimmer Biomet Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2018-RN-00773-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2018-07-09
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    Zimmer biomet, on behalf of lmt surgical, has been made aware via complaint investigation that unexpected lumen may be present in the interface of the handle with the metal chisel. products exhibiting this issue would not be able to be reprocessed with the validated reprocessing cycles.
  • Action
    Zimmer Biomet is advising customers to return all affected product. This item is no longer supplied in Australia.

Device

  • Model / Serial
    Charcot Osteotome ChiselsMultiple Item Numbers and ProductsARTG Number: TBD
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Manufacturer comment
    “The safety of patients has always been, and continues to be, Zimmer Biomet’s top priority and it is our honor to be a leader in this industry for the past 90 years,” Zimmer Biomet told ICIJ in a statement. “We adhere to strict regulatory standards, and work closely with the FDA and all applicable regulatory agencies in each of our regions as part of our commitment to operating a first-rate quality management system across our global manufacturing network. The company added that it is focused on staying at the forefront of innovation and doing right by the millions of patients who rely on the company’s products.
  • Source
    DHTGA