Recall of CEDIA Tobramycin II Assay(used for quantitative determination of tobramycin in human serum or plasma) (an in vitro diagnostic medical device (IVD))

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Thermo Fisher Scientific.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2013-RN-01097-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2013-10-24
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    In-house testing of the cedia tobramycin ii assay (ln: 60169764) shows that the reconstituted reagents do not maintain 30 days reconstituted stability. the instability of the reconstituted reagents may cause higher recovery of samples. hence, lot number 60169764 reagents must be used within 14 days after reconstitution. reconstituted reagents beyond the 14 days should be discontinued and discarded.
  • Action
    Thermo Fisher is advising users to discard the affected lot 14 days after reconstitution. Thermo Fisher is advising users to inform the requesting physician of any change in assay performance and review patient results.