Recall of CARESCAPE Monitor B850 (with software versions 1.0.12 and earlier) and CARESCAPE Monitor B650 (with software version 1.1.12 and earlier)

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by GE Healthcare Australia Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2013-RN-01109-1
  • Event Risk Class
    Class I
  • Event Initiated Date
    2013-10-31
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    Non-invasive blood pressure (nibp) - auto-cycling not starting/stopping, nibp measurement not performed if network time update occurs at same time, infants limits are incorrect. ecg- wrong ecg bandwidth in e-psm(p) or e-(p)re(s)tn module. changing primary ecg lead selection will not change primary measurement in pdm or tram.Central monitor- a temporarily hidden alarm message will not reactivate if different alarm is acknowledged, audible alarm missing after broken network connection, ecg not shown on icentral, standby doesn't end, & incorrect labelling on v leads.Ventilation alarms- incorrect display of airway pressure in bed to bed view, spirometry alarm limits can't be changed.General- alarm off issues in small parameter window, auto-discharge doesn't erase data, invasive pressure lables incorrect.Tram -missing service alarm bed to bed view doesn't show highest st value in s/5 network, missing remote alarm & avoa alarm, monitor disconnect alarms not restored.
  • Action
    GE Healthcare is advising users of work around instructions to follow as an interim measure. GE is providing users with a laminated sheet summarising the work around instructions to keep near the affected units for easy reference until the software update is implemented to correct the issues. This action has been closed-out on 12/02/2016.