Recall of C-series carry bars (used with C300, C450 Manual, C625 ceiling lifts/hoists)

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Rehab Assist Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2016-RN-01200-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2016-09-14
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    Some c-series carry bars are attached to the track using a plastic swivel. if the plastic swivel in these bars becomes damaged there is the potential for it to separate from the carry bar, resulting in a patient fall. rehab assist will supply a free metal insert to be placed inside affected carry bars to prevent separation if the plastic swivel becomes damaged.There have been nine incidents worldwide in which the plastic swivel mechanism has been reported as damaged. no incidents have been reported in australia and there have been no injuries reported as a result of this issue.
  • Action
    Rehab Assist is advising distributors to contact all end users and inspect the swivels to determine if there are any damaged swivels in use. Customers are advised to inspect the swivel for damage or contact their distributor to arrange for a technician to do an inspection. If there is no damage the lift can continue to be used as normal. The metal insert should be installed as soon as practicable. If the swivel appears damaged, do not use the lift until the metal insert has been inserted. Wherever the plastic swivel is damaged, users are also advised to purchase and replace the carry bar as soon as possible. For more details, please see https://www.tga.gov.au/alert/rehab-assist-c-series-ceiling-hoists .

Device

  • Model / Serial
    C-series carry bars (used with C300, C450 Manual, C625 ceiling lifts/hoists)Models Affected: 360741, 360750, 360751, 360755, 360755 Rev B, 360756, 360757 (identifiable by a black plastic swivel/puck)ARTG Number: 169639