Recall of Base array component of the MAKO RIO Robotic Arm

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Stryker Australia Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2015-RN-00873-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2015-09-08
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    The base array and the base array connector is a two piece component on the rio that is used by the rio camera to determine location of the rio relative to the patient. the pin which olds the base array in position and prevents movement has been visually confirmed to have been damaged during assembly.Incidences have been reported where the locking pin on the rio base array may be damaged during assembly onto the rio base array connector.The incidence of this issue is uncommon (only 0.15% of cases over the past 36 months) and, of the reported cases, all were successfully completed with no reported injury to the patient or operating room staff.
  • Action
    Users are requested to complete and return the notification response. A Stryker Representative will be in contact and a new RIO Base Array Connector and Base Array will be installed on the RIO system during regularly scheduled product maintenance. A workaround is also provided for interim use of the system.

Device

  • Model / Serial
    Base array component of the MAKO RIO Robotic ArmCatalogue Numbers:207300 (207956) MAKO RIO Robotic Arm112220 RIO Base Array205143 RIO Base Array ConnectorLot/Serial Numbers: AllARTG Number: 223105
  • Manufacturer

Manufacturer

  • Manufacturer Parent Company (2017)
  • Source
    DHTGA