Recall of AU5800 Clinical Chemistry Analyzer, software versions 4.05 and prior. An in vitro diagnostic medical device (IVD)

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Beckman Coulter Australia Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2012-RN-01255-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2012-12-17
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    Due to a software data storage limitation which only occurs under certain conditions, measurements of the reagent blank (rb) or calibration (cal) sample results are not saved and the software reverts to a previously stored, acceptable rb/cal.
  • Action
    Beckman Coulter is providing updated work instructions to mitigate the issue. A software update will be implemented as a permanent solution.

Device

  • Model / Serial
    AU5800 Clinical Chemistry Analyzer, software versions 4.05 and prior. An in vitro diagnostic medical device (IVD)All serial numbers affectedAll software versions 4.05 and prior are affected.An in vitro diagnostic medical device
  • Manufacturer

Manufacturer