Recall of Artis zee Systems with generator A100Plus or Artis zeego Systems with software version VC21B. (fluoroscopic angiographic x-ray systems)

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Siemens Healthcare Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2016-RN-00375-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2016-03-31
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    The manufacturer, through post market surveillance, has identified the following problems with the affected devices:a) in artis zeego systems, angulations in the vicinity of the c-arm collision area can cause the cable inlet to become clamped and can result in mechanical damage. b) in artis systems with a100plus generators, an attempt to resume operation following detection of a fault (such as a short circuit in the x-ray tube) can result in the failure of a module in the high-voltage generator, which can lead to system non-availability.The reported issues can cause an ongoing procedure to be terminated.
  • Action
    Siemens service organisation will contact the affected customers to arrange for the installation of a software update, which will eliminate the reported problems.

Device

  • Model / Serial
    Artis zee Systems with generator A100Plus or Artis zeego Systems with software version VC21B. (fluoroscopic angiographic x-ray systems) Catalogue/Lot numbers: 10280959, 10094139, 10094135, 10094137, 10094141ARTG Number: 102177