Recall of Artis Q and Q.zen Floorstand and ArtisQ and Q.zen Biplane Floorstand systems with SW version VD10E

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Siemens Healthcare Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2015-RN-00626-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2015-07-14
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    If a position sensor fault in the swivel base axis is not detected by the system software, the c-arm system may move faster than normal when movement is enabled by the operator. the system will not move of its own accord, movement must still be initiated by the operator, however the system can unexpectedly exceed its usual speed, potentially resulting in a collision. there have been no complaints from australian customers regarding this issue.
  • Action
    This issue will be remedied with software update AX037 /15/S (Patch 2). Following the installation of this update, the speed of the C-arm will be limited to a pre-determined level. This action has been closed-out on 12/08/2016.

Device

  • Model / Serial
    Artis Q and Q.zen Floorstand and ArtisQ and Q.zen Biplane Floorstand systems with SW version VD10ECatalogue/Lot number – 10848280, 10848355, 10848282 with SW version VD10EARTG Number: 102177