Recall of Arthrex ITD Articulating Arm 300 + 350MM

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Device Technologies Australia Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2016-RN-01192-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2016-09-12
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    The manufacturer has identified that the rivets of the rotation on the item ‘vesa 75 / 100 adapter connection with rotation’, which is an integral part of the articulating arm 300 + 350mm, may not have been executed in accordance with the relevant standards. the rivets may be weak. in exceptional cases, the connection panel may become loose or detached when subjected to continuous operation. this may potentially endanger the safety of patients, users or third parties. there have been no reported cases of loss or damage in the field related to this issue.
  • Action
    Device Technologies is advising users that a repair kit will be provided to all affected customers to correct the device. This action has been closed-out on 18/05/2017.