Recall of AMH Gen II ELISA (clinical chemistry substrate, Anti-Mullerian Hormone assay). An in vitro diagnostic medical device (IVD).

According to Department of Health, Therapeutic Goods Administration, this recall involved a device in Australia that was produced by Beckman Coulter Australia Pty Ltd.

What is this?

A correction or removal action taken by a manufacturer to address a problem with a medical device. Recalls occur when a medical device is defective, when it could be a risk to health, or when it is both defective and a risk to health.

Learn more about the data here
  • Type of Event
    Recall
  • Event ID
    RC-2012-RN-01227-1
  • Event Risk Class
    Class II
  • Event Initiated Date
    2012-12-05
  • Event Country
  • Event Source
    DHTGA
  • Event Source URL
  • Notes / Alerts
    Australian data is current through July 2018. All of the data comes from the Australian Therapeutic Goods Administration, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Australia.
  • Extra notes in the data
  • Reason
    Beckman coulter has confirmed through customer feedback and internal investigation that some samples diluted with the amh gen ii sample diluent prior to running the assay may generate results that are falsely elevated.
  • Action
    Beckman Coulter is providing updated work around instructions to mitigate the issue and asking users to notify clinicians/research scientists regarding the potentially affected results at the discretion of the Laboratory Director.